The electrochemical method uses the innovative technology of ionization reduction

Operating Instruction

  1. Take off the protective cap from the metal probe
  2. Press OFF/ON button Tester display “0″.
  3. Quick testing: Water the soil with distilled water before testing. Make sure the water can go 10 cm deep at the least. The best humidity is 60%-80%. Insert the metal probe into the soil vertically and clockwise, to about 6-8 cm deep. Make sure the soil have thorough contact with the metal probe. After about 6 seconds, the tester will display the measured value. The tightness of soil will affect the measuring result. Multiple tests at the same and different points should be performed and take the average as the Final result.
    Accurate testing: use a container, mix dried water at the rate of 3:1. Stir and make it into a mud form. Put the metal probe into the mud about 6-8 cm deep. In about 6 seconds, the tester will display the measured value.
  4. Wipe the metal probe with a towel or tissue paper after each test. Make it clean and dry. till the display shows “0″.

Technical Specification (6 Position):

Range : 0-1999 mg/Kg(mg/l)
Resolution : 1mg/kg(mg/l)
Accuracy : ± 0.2 F.S.
Operating Temp : 5-45 °C (41-113°F) /ATC
Calibration : Factory calibrated
Power supply : DC 9V (9V580/6F22Ni-MH rechargeable battery)
Display : LCD with backlight
Dimension : Æ50.5 x 250 mm
Probe length : 140mm
Features : Auto switch off in about 5 minutes